Rapid Temperature Change Test Chamber

Rapid Temperature Change Test Chamber

Advanced control technology
Low energy consumption
Stable and reliable

Send Enquiry

Main Technical Specifications

Select a heating/cooling-rate tab to compare the available models. Each tab lists the verified workspace volume, internal dimensions, external dimensions, viewing-window size and installed power for that rate class.

Temperature RangeHumidity RangePower SupplyRate Test Interval
-70℃ to +150℃20–98% RH380 V+85℃ to -55℃

Performance and selection notes

  • Heating/cooling rates are tested from +85℃ to -55℃ in accordance with GB/T 5170.2-2017.
  • Optional minimum temperature: -75℃; optional minimum humidity: 10% RH.
  • Humidity performance: fluctuation ≤±1.5% RH; uniformity ≤±5% RH; deviation ≤±3% RH above 75% RH and ≤±5% RH at or below 75% RH.
  • Temperature performance: fluctuation ≤±0.5℃; deviation ≤±2.0℃; uniformity ≤2.0℃.
  • Some models are available only in the CTE temperature-only series; custom configurations are available.
ModelVolume (L)Workspace D × W × H (mm)Exterior D × W × H (mm)Viewing Window W × H (mm)Installed Power T / T&H (kW)
CTE/H-SG7015C-05F/W150500×500×6001510×700×(1680+250)300×3509.252 / 9.972
CTE/H-SG7022C-05F/W225500×600×7501550×800×(1830+250)300×35013.152 / 13.272
CTE/H-SG7031C-05F/W315600×700×7501560×900×(1830+250)400×50013.32 / 14.04
CTE/H-SG7040C-05F/W408600×800×8501650×1000×1980400×50016.02 / 16.62
CTE/H-SG7050C-05F/W504700×800×9001650×1000×1980400×50016.02 / 16.62
CTE/H-SG7060C-05F/W600700×900×9501650×1100×2030450×55019.368 / 19.608
CTE/H-SG7080C-05F/W810900×900×10001850×1100×2080450×55019.848 / 20.928
CTE/H-SG70100C-05F/W10001000×1000×10001950×1200×2080450×55019.848 / 20.928
ModelVolume (L)Workspace D × W × H (mm)Exterior D × W × H (mm)Viewing Window W × H (mm)Installed Power T / T&H (kW)
CTE/H-SG7022C-10W225500×600×7501750×800×(1830+250)300×35022.92 / 22.92
CTE/H-SG7031C-10W315600×700×7501850×900×(1830+250)400×50023.688 / 23.688
CTE/H-SG7040C-10W408600×800×8501800×1000×(1930+250)400×50029.388 / 29.388
CTE/H-SG7050C-10W504700×800×9001900×1000×(1980+250)400×50029.388 / 29.388
CTE/H-SG7060C-10W600700×900×9501900×1100×(1980+250)450×55035.256 / 35.256
CTE/H-SG7080C-10W810900×900×10002100×1100×(1990+250)450×55035.736 / 36.408
CTE/H-SG70100C-10W10001000×1000×10002200×1200×(1990+250)450×55035.736 / 36.408
ModelVolume (L)Workspace D × W × H (mm)Exterior D × W × H (mm)Viewing Window W × H (mm)Installed Power T / T&H (kW)
CTE/H-SG7022C-15W225500×600×7501750×800×(1830+250)300×35034.488 / 34.488
CTE/H-SG7031C-15W315600×700×7501850×900×(1830+250)400×50040.368 / 40.368
CTE/H-SG7040C-15W408600×800×8501800×1000×(1930+250)400×50042.048 / 42.048
CTE/H-SG7050C-15W504700×800×9001900×1000×(1980+250)400×50042.048 / 42.048
CTE/H-SG7060C-15W600700×900×9502000×1100×(2080+250)450×55046.248 / 46.248
CTE/H-SG7080C-15W810900×900×10002230×1100×(2130+250)450×55054.12 / 54.12
CTE/H-SG70100C-15W10001000×1000×10002330×1200×(2130+250)450×55054.12 / 53.4
ModelVolume (L)Workspace D × W × H (mm)Exterior D × W × H (mm)Viewing Window W × H (mm)Installed Power T / T&H (kW)
CTE/H-SG7022C-20W225500×600×7501860×800×(1830+250)300×35041.568 / 41.568
CTE/H-SG7031C-20W315600×700×7502050×900×(1880+250)400×50046.248 / 46.248
CTE/H-SG7040C-20W408600×800×8502050×1350×(1980+250)400×50053.64 / 53.64
CTE/H-SG7050C-20W504700×800×9002250×1350×(1980+250)400×50062.28 / 62.28
CTE/H-SG7060C-20W600700×900×9502250×1450×(2080+250)450×55062.28 / 62.28
CTE/H-SG7080C-20W810900×900×10002250×1450×(2130+250)450×55076.68 / 76.68
CTE/H-SG70100C-20W10001000×1000×10002350×1550×(2130+250)450×55094.2 / 94.2
ENGINEERING CONSULTATION

Turn your requirement into a buildable chamber specification.

Share the essentials. Our application engineers will help define a practical, reliable system.

Application and standards reviewPractical chamber configurationInitial response within 1 business day
SCI INSTRUMENTS CO.LTD
info@scicoolings.com
PROJECT ENQUIRY

Request Engineering Consultation

Tell us the essentials. Our application engineers will help complete the specification.

Quick Chamber Selection
Select your test method, control variables and workspace. SCI will use these inputs to recommend the practical configuration.