Two-Zone Basket-Type Thermal Shock Test Chamber

Vertical-Basket Temperature Shock Test Chamber

Advanced control technology
Low energy consumption
Stable and reliable

Send Enquiry

Primary Applications

  • Thermal-shock screening of electronic components, connectors, soldered assemblies and material joints.
  • Evaluation of expansion mismatch, cracking, seal degradation and intermittent electrical faults.
  • Development qualification using repeated hot/cold exposure and defined dwell times.
  • Comparison of design changes under an agreed specimen load and temperature-shock profile.

Product Features

  • Two-zone basket-transfer design moves specimens between separately conditioned hot and cold zones.
  • Hot-zone operating range: +80°C to +180°C; cold-zone operating range: −70°C to +150°C.
  • Reference shock profile: +150°C / −55°C, with 30-minute exposures and model-specific aluminium loads.
  • Basket-transfer time and loaded recovery performance are stated separately for each model.
  • Vertical basket-transfer arrangement, with 49, 100 and 210 L basket models listed below.

Applicable Test Standards

Select the exposure method, then confirm the test profile and specimen load for your application.

Rapid temperature changes using air or a suitable gas; transfer or changeover time and loaded recovery must match the test profile.

Define in your protocol

  • Exposure temperatures, dwell time and cycle count
  • Transfer/changeover time and specimen operating state

Confirm for this chamber

  • Loaded recovery and specimen stabilization
  • Separate zone operating limits from the validated shock profile

Official standard reference ↗

Semiconductor device temperature cycling.

Define in your protocol

  • Selected test condition, cycle count and dwell periods
  • Electrical measurements and device acceptance criteria

Confirm for this chamber

  • Usable working zone and actual specimen temperature
  • Fixtures, loading arrangement and transition performance

Official standard reference ↗

Atmospheric temperature-shock profiles tailored to the specimen and intended environment.

Define in your protocol

  • Applicable revision, procedure and tailored severity
  • Operational state, exposure sequence and functional checks

Confirm for this chamber

  • Achievable transition and loaded temperature response
  • Measurement points, recovery and configuration review

Official standard reference ↗

Related Industry Requirements

Zone operating limits are different from the validated loaded shock profile. Confirm specimen temperature, transfer time and recovery criteria for the intended method. The applicable edition, test severity, load and monitoring requirements are confirmed for the ordered configuration. Standard references do not imply complete product certification.

Technical Specifications

Current standard specifications
Model / platformCTS-DD7505CV-AW/BWCTS-DD7510CV-AW/BWCTS-DD7521CV-AW/BW
Nominal workspaceL49100210
Interior dimensionsW × H × D (mm)400 × 350 × 350500 × 400 × 500650 × 500 × 650
Exterior dimensionsW × H × D (mm)1060 × 1950 × 19501210 × 1980 × 20901360 × 2180 × 2300
Observation windowW × H (mm)250 × 250300 × 300350 × 350
Cable port diametermm5050100
Shelf loadkg/shelf151525
Unit weightkg100012001500
Cold-zone operating range°C−70°C to +150°C
Hot-zone operating range°C+80°C to +180°C
Temperature fluctuation°C≤±0.5°C; empty, steady state
Temperature uniformity°C≤2.0°C; empty, steady state
Temperature deviation°C≤±2.0°C; empty, steady state
Shock load at +150 / −55°Ckg aluminium51020
Shock load at +150 / −65°Ckg aluminium138
Temperature recovery5 min; 30 min dwell in each zone; air-outlet measurement
Basket transfer / air-damper changeovers≤5
Power supplyAC 380 V ±10%, 50 Hz ±1 Hz, 3L+N+PE
Installed powerkW19.72353
Supply circuit breakerA5063125
Cooling-water requirement3/1.6/1 t/h at 29 / 24 / 13°C, water-cooled configuration4/2/1.2 t/h at 29 / 24 / 13°C, water-cooled configuration9/4.5/4 t/h at 29 / 24 / 13°C, water-cooled configuration
Cooling-water connectionDN25DN32DN40
Temperature sensorType-T thermocouple
Communication interfacesOne RS232 interface; one RJ45 Ethernet interface; one USB data interface; RS485 optional

Reference recovery uses +150°C / −55°C, 30-minute exposures and the model-specific non-self-heating aluminium load. Hot-zone and cold-zone limits are conditioning ranges; they are distinct from the reference shock profile.

Recovery is measured at the air outlet after 30-minute exposures at +150°C and −55°C / −65°C, using the corresponding aluminium loads in the table. Shelf capacity, shock load and zone operating range are separate specifications.

Construction and configuration

CompressorImported-brand compressor
Refrigeration componentsSporlan, OLAB, Saginomiya, Danfoss, Emerson or equivalent specified components
Interior material304 stainless steel
Exterior materialElectrostatically powder-coated cold-rolled steel
Observation window configurationOne coated, heated, frost-resistant insulated-glass window with illumination
HeaterGrouped nickel-chromium heating elements with ceramic insulating supports
ControllerSCI TP-V3.0 controller: industrial host, PLC and proprietary control software
Touchscreen12-inch colour touchscreen
Safety protectionEarth-leakage, phase-sequence and overload protection; refrigeration overpressure / overheating protection; fan overload and heater overtemperature protection; specimen overtemperature protection; low-water and humidifier dry-heating protection
Standard accessoriesCasters and levelling feet; one cable port with silicone plug; two specimen shelves; 5 m cable
Software and documentation optionsCustom engineering, remote-control software, third-party calibration and certification reports
Accessory optionsInsulated specimen shelves, custom fixtures, service tools and dry-air system
Installation environmentAmbient temperature 5–35°C; ambient humidity ≤85% RH. Provide ventilation, avoid vibration, electromagnetic interference, dust and corrosive substances. Reserve at least 1 m around enclosed sides and door width + 1 m in front. Drain within 5 m of the refrigeration unit, height ≤30 mm; Ø16 mm quick-connect drain.
ENGINEERING CONSULTATION

Turn your requirement into a buildable chamber specification.

Share the essentials. Our application engineers will help define a practical, reliable system.

Application and standards reviewPractical chamber configurationInitial response within 1 business day
SCI INSTRUMENTS CO.LTD
info@scicoolings.com
PROJECT ENQUIRY

Request Engineering Consultation

Tell us the essentials. Our application engineers will help complete the specification.

Quick Chamber Selection
Select your test method, control variables and workspace. SCI will use these inputs to recommend the practical configuration.