Primary Applications
- Low-pressure exposure of electronic components, assemblies and equipment used or transported at altitude.
- Combined temperature and pressure evaluation within the specified operating envelope.
- Development of storage, transportation and operational test profiles for aerospace and industrial equipment.
- Humidity exposure on HA configurations when included in the approved test arrangement.
Product Features
- Standard temperature range: −70°C to +150°C; pressure range: atmospheric pressure to 0.5 kPa.
- 1000, 225 and 504 L workspace options are identified by the actual model code.
- Pressure control, pump-down, repressurization and segmented pressure-deviation limits are listed in the model matrix.
- Temperature operation below 5 kPa requires a separately agreed configuration.
- HA humidity capability is optional; simultaneous humidity and reduced-pressure operation is not part of the standard configuration.
Applicable Test Standards
Select the exposure method, then confirm the test profile and specimen load for your application.
Reduced-air-pressure exposure for transport, storage or operation.
Define in your protocol
- Pressure level, duration and rate of pressure change
- Operational state, functional checks and recovery
Confirm for this chamber
- Chamber pressure envelope, pump-down and repressurization
- Pressure measurement, feedthroughs and specimen heat release
Combined temperature and reduced-pressure testing; simultaneous humidity requires a separately agreed configuration and operating envelope.
Define in your protocol
- Combined exposure sequence, pressure and temperature levels
- Specimen operating state and functional measurements
Confirm for this chamber
- Approved combined temperature/pressure operating envelope
- Humidity only where specifically engineered and agreed
Cold exposure of components and equipment.
Define in your protocol
- Lower temperature, tolerances and exposure duration
- Specimen operating state, stabilization and recovery
Confirm for this chamber
- Minimum temperature for the selected model and load
- Usable test space, sensors and loaded uniformity
Additional methods: IEC 60068-2-2:2025
Related Industry Requirements
MIL-STD-810H + Change 1
Environmentally tailored profiles
Select the method and severities for the intended service environment. Confirm the applicable revision, specimen operating state, monitoring locations and loaded test performance.
ISO 16750-4:2023
Climatic loads for vehicle electronics
Select the climatic program for the mounting location and intended operating state. Confirm specimen supply, heat dissipation, tolerances and functional checks.
IEC 60068 series / laboratory protocols
Document the actual exposure method
Define the relevant method or laboratory procedure, specimen load, measurement locations, calibration and repeatability criteria. Equipment selection follows the required profile and documented uncertainty.
Select test severities against the approved temperature/pressure envelope, specimen heat dissipation and pressure-change requirements. The applicable edition, test severity, load and monitoring requirements are confirmed for the ordered configuration. Standard references do not imply complete product certification.
Technical Specifications
| Current standard specifications | ||||
|---|---|---|---|---|
| Model / platform | CTA/HA-SG7022C-02F/W | CTA/HA-SG7050C-02F/W | CTA/HA-SG70100C-02F/W | |
| Nominal workspace | L | 225 | 504 | 1000 |
| Interior dimensions | W × H × D (mm) | 600 × 750 × 500 | 800 × 900 × 700 | 1000 × 1000 × 1000 |
| Exterior dimensions | W × H × D (mm) | 1000 × 2450 × 2200 | 1200 × 2550 × 2400 | 1400 × 2665 × 2700 |
| Observation window | W × H (mm) | Ø200 mm | ||
| Cable port diameter | mm | 100 | ||
| Shelf load | kg/shelf | 15 | 25 | 25 |
| Unit weight | kg | 1300 | 1500 | 2300 |
| Temperature range | °C | −70°C to +150°C | ||
| Temperature fluctuation | °C | ≤±0.5°C; empty, steady state | ||
| Temperature uniformity | °C | ≤2.0°C; empty, steady state | ||
| Temperature deviation | °C | ≤±2.0°C; empty, steady state | ||
| Average heating / cooling rate | °C/min | ≥2.5°C/min (+85°C → −40°C); ≥2°C/min (+85°C → −55°C); average; 25 kg aluminium; air outlet | ||
| Humidity range and operating temperature | % RH | 20–98% RH at +20°C to +85°C; CTHA models only | ||
| Humidity fluctuation | % RH | ≤±3.0% RH; empty, steady state | ||
| Humidity uniformity | % RH | ≤3.0% RH; empty, steady state | ||
| Humidity deviation | % RH | ≤±3.0% RH above 75% RH; ≤±5.0% RH at or below 75% RH; empty, steady state | ||
| Power supply | AC 380 V ±10%, 50 Hz ±1 Hz, 3L+N+PE | |||
| Installed power | kW | 18 | 26 | 38 |
| Supply circuit breaker | A | 40 | 63 | 80 |
| Cooling-water requirement | 3.2/1.6/1.1 t/h at 29 / 24 / 13°C, water-cooled configuration | 6/3/2 t/h at 29 / 24 / 13°C, water-cooled configuration | 8/4/2.7 t/h at 29 / 24 / 13°C, water-cooled configuration | |
| Cooling-water connection | DN25 | DN32 | DN40 | |
| Dry-air supply | 0.6–0.8 MPa; dew point −60°C; ≥200 L/min; Ø10 mm connector | |||
| Pressure capability | Atmospheric pressure to 0.5 kPa; external-pressure method as standard | |||
| Humidification-water resistivity | ≥500 Ω·m | |||
| Temperature sensor | PT100 | |||
The CTA configuration is temperature/pressure controlled; CTHA adds humidity within its operating envelope. Standard equipment does not control humidity and reduced pressure simultaneously. Simultaneous pressure–humidity control requires a separately specified internal-pressure configuration. Temperature control below 5 kPa is a custom requirement. Temperature performance is specified at atmospheric pressure and 25°C ambient.
CTA provides temperature / low-pressure control; CTHA adds humidity. Humidity values apply to CTHA. Standard construction uses external pressure. Simultaneous pressure / humidity control and temperature control at or below 5 kPa require a specified custom configuration. Pressure deviation is ±2 kPa from atmospheric pressure to 40 kPa, ±5% from 40 to 4 kPa and ±0.1 kPa from 4 to 1 kPa. Pump-down to 0.5 kPa and repressurization take no more than 60 and 30 minutes respectively. These new -02F/W references are quoted separately from earlier -01W configurations.
Construction and configuration
| Compressor | Imported-brand compressor |
|---|---|
| Refrigeration components | Sporlan, OLAB, Saginomiya, Danfoss, Emerson or equivalent specified components |
| Interior material | 304 stainless steel |
| Exterior material | Electrostatically powder-coated cold-rolled steel |
| Observation window configuration | One coated, heated, frost-resistant insulated-glass window with illumination |
| Heater | Grouped nickel-chromium heating elements with ceramic insulating supports |
| Controller | SCI TP-V3.0 controller: industrial host, PLC and proprietary control software |
| Humidity sensor | Capacitive humidity sensor |
| Touchscreen | 12-inch colour touchscreen |
| Safety protection | Earth-leakage, phase-sequence and overload protection; refrigeration overpressure / overheating protection; fan overload and heater overtemperature protection; specimen overtemperature protection; low-water and humidifier dry-heating protection |
| Standard accessories | Casters and levelling feet; two cable ports and silicone plugs; two specimen shelves; 5 m cable; two blank flanges |
| Software and documentation options | Custom engineering, remote-control software, third-party calibration and certification reports |
| Accessory options | Connector flanges, terminals, insulated shelves, custom fixtures, service tools and dry-air system |
| Installation environment | Ambient temperature 5–35°C; ambient humidity ≤85% RH. Provide ventilation, avoid vibration, electromagnetic interference, dust and corrosive substances. Reserve at least 1 m around enclosed sides and door width + 1 m in front. Drain within 5 m of the refrigeration unit, height ≤30 mm; Ø16 mm quick-connect drain. |
Earlier model configurations
Select an existing model reference below when enquiring about a previous configuration. Availability and the ordered configuration are confirmed in the quotation.
| Main Technical Specifications | ||||
|---|---|---|---|---|
| Model | Model | CTA/HA-SG7022C-01W | CTA/HA-SG7050C-01W | CTA/HA-SG70100C-01W |
| Interior volume | L | 225 | 504 | 1000 |
| Interior dimensions | W × H × D (mm) | 600 × 750 × 500 | 800 × 900 × 700 | 1000 × 1000 × 1000 |
| Exterior dimensions | W × H × D (mm) | 1000 × 2450 × 2200 | 1200 × 2550 × 2400 | 1400 × 2665 × 2700 |
| Observation window | W × H (mm) | φ200mm | φ200mm | φ200mm |
| Cable port diameter | mm | 100 | 100 | 100 |
| Shelf load | kg/shelf | 15 | 25 | 25 |
| Unit weight | kg | 1300 | 1500 | 2300 |
| Temperature range | °C | -70 to +150 | -70 to +150 | -70 to +150 |
| Temperature fluctuation | °C | ≤±0.5 (empty, steady state) | ≤±0.5 (empty, steady state) | ≤±0.5 (empty, steady state) |
| Temperature uniformity | °C | ≤2.0 (empty, steady state) | ≤2.0 (empty, steady state) | ≤2.0 (empty, steady state) |
| Temperature deviation | °C | ≤±2.0 (empty, steady state) | ≤±2.0 (empty, steady state) | ≤±2.0 (empty, steady state) |
| Temperature-change rate | °C/min | ≥2.5, average; 25 kg non-self-heating aluminium load; −40°C ↔ +85°C; air-outlet measurement | ≥2.5, average; 25 kg non-self-heating aluminium load; −40°C ↔ +85°C; air-outlet measurement | ≥2.5, average; 25 kg non-self-heating aluminium load; −40°C ↔ +85°C; air-outlet measurement |
| Temperature-change rate | °C/min | ≥2.0, average; 25 kg non-self-heating aluminium load; −55°C ↔ +85°C; air-outlet measurement | ≥2.0, average; 25 kg non-self-heating aluminium load; −55°C ↔ +85°C; air-outlet measurement | ≥2.0, average; 25 kg non-self-heating aluminium load; −55°C ↔ +85°C; air-outlet measurement |
| Humidity range | % RH | 20 to 98 | 20 to 98 | 20 to 98 |
| Humidity fluctuation | % RH | ≤±3.0 (empty, steady state) | ≤±3.0 (empty, steady state) | ≤±3.0 (empty, steady state) |
| Humidity uniformity | % RH | ≤3.0 (empty, steady state) | ≤3.0 (empty, steady state) | ≤3.0 (empty, steady state) |
| Humidity deviation | % RH | ≤±3.0 above 75% RH (empty, steady state); ≤±5.0 at or below 75% RH (empty, steady state) | ≤±3.0 above 75% RH (empty, steady state); ≤±5.0 at or below 75% RH (empty, steady state) | ≤±3.0 above 75% RH (empty, steady state); ≤±5.0 at or below 75% RH (empty, steady state) |
| Pressure range | atmospheric pressure to 0.5kPa | atmospheric pressure to 0.5kPa | atmospheric pressure to 0.5kPa | |
| Repressurization time | ≤30min (0.5kPa → atmospheric pressure) | ≤30min (0.5kPa → atmospheric pressure) | ≤30min (0.5kPa → atmospheric pressure) | |
| Pump-down time | ≤60min (atmospheric pressure → 0.5kPa) | ≤60min (atmospheric pressure → 0.5kPa) | ≤60min (atmospheric pressure → 0.5kPa) | |
| Pressure deviation | ±2 kPa from atmospheric pressure to 40 kPa; ±5% from 40 to 4 kPa; ±0.1 kPa from 4 to 1 kPa | ±2 kPa from atmospheric pressure to 40 kPa; ±5% from 40 to 4 kPa; ±0.1 kPa from 4 to 1 kPa | ±2 kPa from atmospheric pressure to 40 kPa; ±5% from 40 to 4 kPa; ±0.1 kPa from 4 to 1 kPa | |
| Refrigeration system | Mechanical cascade refrigeration | Mechanical cascade refrigeration | Mechanical cascade refrigeration | |
| Cooling method | Water-cooled | Water-cooled | Water-cooled | |
| Compressor | Imported-brand compressor | Imported-brand compressor | Imported-brand compressor | |
| Refrigerant | R507/R404A+R23/R508B | R507/R404A+R23/R508B | R507/R404A+R23/R508B | |
| Refrigeration components | Sporlan, OLAB, Saginomiya, Danfoss, Emerson or equivalent specified components | Sporlan, OLAB, Saginomiya, Danfoss, Emerson or equivalent specified components | Sporlan, OLAB, Saginomiya, Danfoss, Emerson or equivalent specified components | |
| Power supply | AC 380±10%V; 50±1Hz; 3L+N+PE | AC 380±10%V; 50±1Hz; 3L+N+PE | AC 380±10%V; 50±1Hz; 3L+N+PE | |
| Installed power | kW | 18 | 26 | 38 |
| Supply circuit breaker | A | 40 | 63 | 80 |
| Interior material | 304 stainless steel | 304 stainless steel | 304 stainless steel | |
| Exterior material | Electrostatically powder-coated cold-rolled steel | Electrostatically powder-coated cold-rolled steel | Electrostatically powder-coated cold-rolled steel | |
| Insulation | Composite polyurethane-foam and glass-fibre insulation | Composite polyurethane-foam and glass-fibre insulation | Composite polyurethane-foam and glass-fibre insulation | |
| Observation window configuration | One coated, heated, frost-resistant insulated-glass window with illumination | One coated, heated, frost-resistant insulated-glass window with illumination | One coated, heated, frost-resistant insulated-glass window with illumination | |
| Heater | Grouped nickel-chromium heating elements with ceramic insulating supports | Grouped nickel-chromium heating elements with ceramic insulating supports | Grouped nickel-chromium heating elements with ceramic insulating supports | |
| Controller | SCI TP-V3.0 controller: industrial host, PLC and proprietary control software | SCI TP-V3.0 controller: industrial host, PLC and proprietary control software | SCI TP-V3.0 controller: industrial host, PLC and proprietary control software | |
| Temperature sensor | PT100 | PT100 | PT100 | |
| Humidity sensor | Capacitive humidity sensor | Capacitive humidity sensor | Capacitive humidity sensor | |
| Pressure sensor | Electronic pressure sensor | Electronic pressure sensor | Electronic pressure sensor | |
| Touchscreen | 12-inch colour touchscreen | 12-inch colour touchscreen | 12-inch colour touchscreen | |
| Communication interfaces | RS232/RS485, RJ45 Ethernet and USB data interfaces | RS232/RS485, RJ45 Ethernet and USB data interfaces | RS232/RS485, RJ45 Ethernet and USB data interfaces | |
| Safety protection | Earth-leakage, phase-sequence and overload protection; refrigeration overpressure / overheating protection; fan overload and heater overtemperature protection; specimen overtemperature protection; low-water and humidifier dry-heating protection | Earth-leakage, phase-sequence and overload protection; refrigeration overpressure / overheating protection; fan overload and heater overtemperature protection; specimen overtemperature protection; low-water and humidifier dry-heating protection | Earth-leakage, phase-sequence and overload protection; refrigeration overpressure / overheating protection; fan overload and heater overtemperature protection; specimen overtemperature protection; low-water and humidifier dry-heating protection | |
| Standard accessories | Casters and levelling feet; two cable ports and silicone plugs; two specimen shelves; 5 m cable; two blank flanges | Casters and levelling feet; two cable ports and silicone plugs; two specimen shelves; 5 m cable; two blank flanges | Casters and levelling feet; two cable ports and silicone plugs; two specimen shelves; 5 m cable; two blank flanges | |
| Software and documentation options | Custom engineering, remote-control software, third-party calibration and certification reports | Custom engineering, remote-control software, third-party calibration and certification reports | Custom engineering, remote-control software, third-party calibration and certification reports | |
| Accessory options | Connector flanges, terminals, insulated shelves, custom fixtures, service tools and dry-air system | Connector flanges, terminals, insulated shelves, custom fixtures, service tools and dry-air system | Connector flanges, terminals, insulated shelves, custom fixtures, service tools and dry-air system | |
| Cooling-water flow at 29 / 24 / 13°C | 3.2/1.6/1.1 | 6/3/2 | 8/4/2.7 | |
| Cooling-water connection | 25 | 32 | 40 | |
| Humidification water | Purified or distilled water; pH 6–8; pressure 1–4 bar; water temperature 5–45°C. Water resistivity requirement to be specified in the agreed utility specification. | Purified or distilled water; pH 6–8; pressure 1–4 bar; water temperature 5–45°C. Water resistivity requirement to be specified in the agreed utility specification. | Purified or distilled water; pH 6–8; pressure 1–4 bar; water temperature 5–45°C. Water resistivity requirement to be specified in the agreed utility specification. | |
| Dry-air supply | Supply pressure 0.4–0.8 MPa; dew point −60°C; flow ≥100 L/min; Ø10 mm quick connector. Customer provides the supply piping. | Supply pressure 0.4–0.8 MPa; dew point −60°C; flow ≥100 L/min; Ø10 mm quick connector. Customer provides the supply piping. | Supply pressure 0.4–0.8 MPa; dew point −60°C; flow ≥100 L/min; Ø10 mm quick connector. Customer provides the supply piping. | |
| Installation environment | Ambient temperature 5–35°C; ambient humidity ≤85% RH. Provide ventilation, avoid vibration, electromagnetic interference, dust and corrosive substances. Reserve at least 1 m around enclosed sides and door width + 1 m in front. Drain within 5 m of the refrigeration unit, height ≤30 mm; Ø16 mm quick-connect drain. | Ambient temperature 5–35°C; ambient humidity ≤85% RH. Provide ventilation, avoid vibration, electromagnetic interference, dust and corrosive substances. Reserve at least 1 m around enclosed sides and door width + 1 m in front. Drain within 5 m of the refrigeration unit, height ≤30 mm; Ø16 mm quick-connect drain. | Ambient temperature 5–35°C; ambient humidity ≤85% RH. Provide ventilation, avoid vibration, electromagnetic interference, dust and corrosive substances. Reserve at least 1 m around enclosed sides and door width + 1 m in front. Drain within 5 m of the refrigeration unit, height ≤30 mm; Ø16 mm quick-connect drain. | |