SCI

Low-Pressure Test Chamber

CTA/HA-SG70 Series Low-Pressure Test Chambers

225, 504 and 1000 L standard models for temperature and reduced-pressure testing, with optional humidity configuration.

Send Enquiry

Primary Applications

  • Low-pressure exposure of electronic components, assemblies and equipment used or transported at altitude.
  • Combined temperature and pressure evaluation within the specified operating envelope.
  • Development of storage, transportation and operational test profiles for aerospace and industrial equipment.
  • Humidity exposure on HA configurations when included in the approved test arrangement.

Product Features

  • Standard temperature range: −70°C to +150°C; pressure range: atmospheric pressure to 0.5 kPa.
  • 1000, 225 and 504 L workspace options are identified by the actual model code.
  • Pressure control, pump-down, repressurization and segmented pressure-deviation limits are listed in the model matrix.
  • Temperature operation below 5 kPa requires a separately agreed configuration.
  • HA humidity capability is optional; simultaneous humidity and reduced-pressure operation is not part of the standard configuration.

Applicable Test Standards

Select the exposure method, then confirm the test profile and specimen load for your application.

Reduced-air-pressure exposure for transport, storage or operation.

Define in your protocol

  • Pressure level, duration and rate of pressure change
  • Operational state, functional checks and recovery

Confirm for this chamber

  • Chamber pressure envelope, pump-down and repressurization
  • Pressure measurement, feedthroughs and specimen heat release

Official standard reference ↗

Combined temperature and reduced-pressure testing; simultaneous humidity requires a separately agreed configuration and operating envelope.

Define in your protocol

  • Combined exposure sequence, pressure and temperature levels
  • Specimen operating state and functional measurements

Confirm for this chamber

  • Approved combined temperature/pressure operating envelope
  • Humidity only where specifically engineered and agreed

Official standard reference ↗

Cold exposure of components and equipment.

Define in your protocol

  • Lower temperature, tolerances and exposure duration
  • Specimen operating state, stabilization and recovery

Confirm for this chamber

  • Minimum temperature for the selected model and load
  • Usable test space, sensors and loaded uniformity

Official standard reference ↗

Additional methods: IEC 60068-2-2:2025

Related Industry Requirements

Select test severities against the approved temperature/pressure envelope, specimen heat dissipation and pressure-change requirements. The applicable edition, test severity, load and monitoring requirements are confirmed for the ordered configuration. Standard references do not imply complete product certification.

Technical Specifications

Current standard specifications
Model / platformCTA/HA-SG7022C-02F/WCTA/HA-SG7050C-02F/WCTA/HA-SG70100C-02F/W
Nominal workspaceL2255041000
Interior dimensionsW × H × D (mm)600 × 750 × 500800 × 900 × 7001000 × 1000 × 1000
Exterior dimensionsW × H × D (mm)1000 × 2450 × 22001200 × 2550 × 24001400 × 2665 × 2700
Observation windowW × H (mm)Ø200 mm
Cable port diametermm100
Shelf loadkg/shelf152525
Unit weightkg130015002300
Temperature range°C−70°C to +150°C
Temperature fluctuation°C≤±0.5°C; empty, steady state
Temperature uniformity°C≤2.0°C; empty, steady state
Temperature deviation°C≤±2.0°C; empty, steady state
Average heating / cooling rate°C/min≥2.5°C/min (+85°C → −40°C); ≥2°C/min (+85°C → −55°C); average; 25 kg aluminium; air outlet
Humidity range and operating temperature% RH20–98% RH at +20°C to +85°C; CTHA models only
Humidity fluctuation% RH≤±3.0% RH; empty, steady state
Humidity uniformity% RH≤3.0% RH; empty, steady state
Humidity deviation% RH≤±3.0% RH above 75% RH; ≤±5.0% RH at or below 75% RH; empty, steady state
Power supplyAC 380 V ±10%, 50 Hz ±1 Hz, 3L+N+PE
Installed powerkW182638
Supply circuit breakerA406380
Cooling-water requirement3.2/1.6/1.1 t/h at 29 / 24 / 13°C, water-cooled configuration6/3/2 t/h at 29 / 24 / 13°C, water-cooled configuration8/4/2.7 t/h at 29 / 24 / 13°C, water-cooled configuration
Cooling-water connectionDN25DN32DN40
Dry-air supply0.6–0.8 MPa; dew point −60°C; ≥200 L/min; Ø10 mm connector
Pressure capabilityAtmospheric pressure to 0.5 kPa; external-pressure method as standard
Humidification-water resistivity≥500 Ω·m
Temperature sensorPT100

The CTA configuration is temperature/pressure controlled; CTHA adds humidity within its operating envelope. Standard equipment does not control humidity and reduced pressure simultaneously. Simultaneous pressure–humidity control requires a separately specified internal-pressure configuration. Temperature control below 5 kPa is a custom requirement. Temperature performance is specified at atmospheric pressure and 25°C ambient.

CTA provides temperature / low-pressure control; CTHA adds humidity. Humidity values apply to CTHA. Standard construction uses external pressure. Simultaneous pressure / humidity control and temperature control at or below 5 kPa require a specified custom configuration. Pressure deviation is ±2 kPa from atmospheric pressure to 40 kPa, ±5% from 40 to 4 kPa and ±0.1 kPa from 4 to 1 kPa. Pump-down to 0.5 kPa and repressurization take no more than 60 and 30 minutes respectively. These new -02F/W references are quoted separately from earlier -01W configurations.

Construction and configuration

CompressorImported-brand compressor
Refrigeration componentsSporlan, OLAB, Saginomiya, Danfoss, Emerson or equivalent specified components
Interior material304 stainless steel
Exterior materialElectrostatically powder-coated cold-rolled steel
Observation window configurationOne coated, heated, frost-resistant insulated-glass window with illumination
HeaterGrouped nickel-chromium heating elements with ceramic insulating supports
ControllerSCI TP-V3.0 controller: industrial host, PLC and proprietary control software
Humidity sensorCapacitive humidity sensor
Touchscreen12-inch colour touchscreen
Safety protectionEarth-leakage, phase-sequence and overload protection; refrigeration overpressure / overheating protection; fan overload and heater overtemperature protection; specimen overtemperature protection; low-water and humidifier dry-heating protection
Standard accessoriesCasters and levelling feet; two cable ports and silicone plugs; two specimen shelves; 5 m cable; two blank flanges
Software and documentation optionsCustom engineering, remote-control software, third-party calibration and certification reports
Accessory optionsConnector flanges, terminals, insulated shelves, custom fixtures, service tools and dry-air system
Installation environmentAmbient temperature 5–35°C; ambient humidity ≤85% RH. Provide ventilation, avoid vibration, electromagnetic interference, dust and corrosive substances. Reserve at least 1 m around enclosed sides and door width + 1 m in front. Drain within 5 m of the refrigeration unit, height ≤30 mm; Ø16 mm quick-connect drain.
Earlier model configurations

Select an existing model reference below when enquiring about a previous configuration. Availability and the ordered configuration are confirmed in the quotation.

Main Technical Specifications
ModelModelCTA/HA-SG7022C-01WCTA/HA-SG7050C-01WCTA/HA-SG70100C-01W
Interior volumeL2255041000
Interior dimensionsW × H × D (mm)600 × 750 × 500800 × 900 × 7001000 × 1000 × 1000
Exterior dimensionsW × H × D (mm)1000 × 2450 × 22001200 × 2550 × 24001400 × 2665 × 2700
Observation windowW × H (mm)φ200mmφ200mmφ200mm
Cable port diametermm100100100
Shelf loadkg/shelf152525
Unit weightkg130015002300
Temperature range°C-70 to +150-70 to +150-70 to +150
Temperature fluctuation°C≤±0.5 (empty, steady state)≤±0.5 (empty, steady state)≤±0.5 (empty, steady state)
Temperature uniformity°C≤2.0 (empty, steady state)≤2.0 (empty, steady state)≤2.0 (empty, steady state)
Temperature deviation°C≤±2.0 (empty, steady state)≤±2.0 (empty, steady state)≤±2.0 (empty, steady state)
Temperature-change rate°C/min≥2.5, average; 25 kg non-self-heating aluminium load; −40°C ↔ +85°C; air-outlet measurement≥2.5, average; 25 kg non-self-heating aluminium load; −40°C ↔ +85°C; air-outlet measurement≥2.5, average; 25 kg non-self-heating aluminium load; −40°C ↔ +85°C; air-outlet measurement
Temperature-change rate°C/min≥2.0, average; 25 kg non-self-heating aluminium load; −55°C ↔ +85°C; air-outlet measurement≥2.0, average; 25 kg non-self-heating aluminium load; −55°C ↔ +85°C; air-outlet measurement≥2.0, average; 25 kg non-self-heating aluminium load; −55°C ↔ +85°C; air-outlet measurement
Humidity range% RH20 to 9820 to 9820 to 98
Humidity fluctuation% RH≤±3.0 (empty, steady state)≤±3.0 (empty, steady state)≤±3.0 (empty, steady state)
Humidity uniformity% RH≤3.0 (empty, steady state)≤3.0 (empty, steady state)≤3.0 (empty, steady state)
Humidity deviation% RH≤±3.0 above 75% RH (empty, steady state); ≤±5.0 at or below 75% RH (empty, steady state)≤±3.0 above 75% RH (empty, steady state); ≤±5.0 at or below 75% RH (empty, steady state)≤±3.0 above 75% RH (empty, steady state); ≤±5.0 at or below 75% RH (empty, steady state)
Pressure rangeatmospheric pressure to 0.5kPaatmospheric pressure to 0.5kPaatmospheric pressure to 0.5kPa
Repressurization time≤30min (0.5kPa → atmospheric pressure)≤30min (0.5kPa → atmospheric pressure)≤30min (0.5kPa → atmospheric pressure)
Pump-down time≤60min (atmospheric pressure → 0.5kPa)≤60min (atmospheric pressure → 0.5kPa)≤60min (atmospheric pressure → 0.5kPa)
Pressure deviation±2 kPa from atmospheric pressure to 40 kPa; ±5% from 40 to 4 kPa; ±0.1 kPa from 4 to 1 kPa±2 kPa from atmospheric pressure to 40 kPa; ±5% from 40 to 4 kPa; ±0.1 kPa from 4 to 1 kPa±2 kPa from atmospheric pressure to 40 kPa; ±5% from 40 to 4 kPa; ±0.1 kPa from 4 to 1 kPa
Refrigeration systemMechanical cascade refrigerationMechanical cascade refrigerationMechanical cascade refrigeration
Cooling methodWater-cooledWater-cooledWater-cooled
CompressorImported-brand compressorImported-brand compressorImported-brand compressor
RefrigerantR507/R404A+R23/R508BR507/R404A+R23/R508BR507/R404A+R23/R508B
Refrigeration componentsSporlan, OLAB, Saginomiya, Danfoss, Emerson or equivalent specified componentsSporlan, OLAB, Saginomiya, Danfoss, Emerson or equivalent specified componentsSporlan, OLAB, Saginomiya, Danfoss, Emerson or equivalent specified components
Power supplyAC 380±10%V; 50±1Hz; 3L+N+PEAC 380±10%V; 50±1Hz; 3L+N+PEAC 380±10%V; 50±1Hz; 3L+N+PE
Installed powerkW182638
Supply circuit breakerA406380
Interior material304 stainless steel304 stainless steel304 stainless steel
Exterior materialElectrostatically powder-coated cold-rolled steelElectrostatically powder-coated cold-rolled steelElectrostatically powder-coated cold-rolled steel
InsulationComposite polyurethane-foam and glass-fibre insulationComposite polyurethane-foam and glass-fibre insulationComposite polyurethane-foam and glass-fibre insulation
Observation window configurationOne coated, heated, frost-resistant insulated-glass window with illuminationOne coated, heated, frost-resistant insulated-glass window with illuminationOne coated, heated, frost-resistant insulated-glass window with illumination
HeaterGrouped nickel-chromium heating elements with ceramic insulating supportsGrouped nickel-chromium heating elements with ceramic insulating supportsGrouped nickel-chromium heating elements with ceramic insulating supports
ControllerSCI TP-V3.0 controller: industrial host, PLC and proprietary control softwareSCI TP-V3.0 controller: industrial host, PLC and proprietary control softwareSCI TP-V3.0 controller: industrial host, PLC and proprietary control software
Temperature sensorPT100PT100PT100
Humidity sensorCapacitive humidity sensorCapacitive humidity sensorCapacitive humidity sensor
Pressure sensorElectronic pressure sensorElectronic pressure sensorElectronic pressure sensor
Touchscreen12-inch colour touchscreen12-inch colour touchscreen12-inch colour touchscreen
Communication interfacesRS232/RS485, RJ45 Ethernet and USB data interfacesRS232/RS485, RJ45 Ethernet and USB data interfacesRS232/RS485, RJ45 Ethernet and USB data interfaces
Safety protectionEarth-leakage, phase-sequence and overload protection; refrigeration overpressure / overheating protection; fan overload and heater overtemperature protection; specimen overtemperature protection; low-water and humidifier dry-heating protectionEarth-leakage, phase-sequence and overload protection; refrigeration overpressure / overheating protection; fan overload and heater overtemperature protection; specimen overtemperature protection; low-water and humidifier dry-heating protectionEarth-leakage, phase-sequence and overload protection; refrigeration overpressure / overheating protection; fan overload and heater overtemperature protection; specimen overtemperature protection; low-water and humidifier dry-heating protection
Standard accessoriesCasters and levelling feet; two cable ports and silicone plugs; two specimen shelves; 5 m cable; two blank flangesCasters and levelling feet; two cable ports and silicone plugs; two specimen shelves; 5 m cable; two blank flangesCasters and levelling feet; two cable ports and silicone plugs; two specimen shelves; 5 m cable; two blank flanges
Software and documentation optionsCustom engineering, remote-control software, third-party calibration and certification reportsCustom engineering, remote-control software, third-party calibration and certification reportsCustom engineering, remote-control software, third-party calibration and certification reports
Accessory optionsConnector flanges, terminals, insulated shelves, custom fixtures, service tools and dry-air systemConnector flanges, terminals, insulated shelves, custom fixtures, service tools and dry-air systemConnector flanges, terminals, insulated shelves, custom fixtures, service tools and dry-air system
Cooling-water flow at 29 / 24 / 13°C3.2/1.6/1.16/3/28/4/2.7
Cooling-water connection253240
Humidification waterPurified or distilled water; pH 6–8; pressure 1–4 bar; water temperature 5–45°C. Water resistivity requirement to be specified in the agreed utility specification.Purified or distilled water; pH 6–8; pressure 1–4 bar; water temperature 5–45°C. Water resistivity requirement to be specified in the agreed utility specification.Purified or distilled water; pH 6–8; pressure 1–4 bar; water temperature 5–45°C. Water resistivity requirement to be specified in the agreed utility specification.
Dry-air supplySupply pressure 0.4–0.8 MPa; dew point −60°C; flow ≥100 L/min; Ø10 mm quick connector. Customer provides the supply piping.Supply pressure 0.4–0.8 MPa; dew point −60°C; flow ≥100 L/min; Ø10 mm quick connector. Customer provides the supply piping.Supply pressure 0.4–0.8 MPa; dew point −60°C; flow ≥100 L/min; Ø10 mm quick connector. Customer provides the supply piping.
Installation environmentAmbient temperature 5–35°C; ambient humidity ≤85% RH. Provide ventilation, avoid vibration, electromagnetic interference, dust and corrosive substances. Reserve at least 1 m around enclosed sides and door width + 1 m in front. Drain within 5 m of the refrigeration unit, height ≤30 mm; Ø16 mm quick-connect drain.Ambient temperature 5–35°C; ambient humidity ≤85% RH. Provide ventilation, avoid vibration, electromagnetic interference, dust and corrosive substances. Reserve at least 1 m around enclosed sides and door width + 1 m in front. Drain within 5 m of the refrigeration unit, height ≤30 mm; Ø16 mm quick-connect drain.Ambient temperature 5–35°C; ambient humidity ≤85% RH. Provide ventilation, avoid vibration, electromagnetic interference, dust and corrosive substances. Reserve at least 1 m around enclosed sides and door width + 1 m in front. Drain within 5 m of the refrigeration unit, height ≤30 mm; Ø16 mm quick-connect drain.
ENGINEERING CONSULTATION

Turn your requirement into a buildable chamber specification.

Share the essentials. Our application engineers will help define a practical, reliable system.

Application and standards reviewPractical chamber configurationInitial response within 1 business day
SCI INSTRUMENTS CO.LTD