Turntable-Compatible Test Chamber

CTT-SC Single-Axis Turntable Test Chamber

Turntable integration: extensive environmental-test integration experience with technical-team support.
Complete range: configurations for single-, dual- and three-axis turntables in multiple sizes.

Send Enquiry

Primary Applications

  • Temperature exposure during single-axis turntable testing of inertial sensors and navigation assemblies.
  • Evaluation of temperature-dependent measurement drift, alignment and functional stability.
  • Conditioned static measurements and calibration using the specified external instruments.
  • Integrated motion and climatic test development for aerospace, electronics and precision equipment.

Product Features

  • Chamber interface engineered around a single-axis turntable, specimen fixture and service connections.
  • Workspace geometry, access and clearances are selected for the actual turntable installation.
  • Temperature range, rate and load conditions are stated for the listed chamber model.
  • Programmable temperature control with communication interfaces for system integration.
  • Motion performance and measurement uncertainty are specified for the complete turntable and instrumentation system.

Applicable Test Standards

Select the exposure method, then confirm the test profile and specimen load for your application.

Cold exposure of components and equipment.

Define in your protocol

  • Lower temperature, tolerances and exposure duration
  • Specimen operating state, stabilization and recovery

Confirm for this chamber

  • Minimum temperature for the selected model and load
  • Usable test space, sensors and loaded uniformity

Official standard reference ↗

Dry-heat exposure of components and equipment.

Define in your protocol

  • Upper temperature, exposure duration and test sequence
  • Specimen heat dissipation, operating state and recovery

Confirm for this chamber

  • Maximum temperature and actual specimen heat release
  • Loaded uniformity, airflow and independent monitoring

Official standard reference ↗

Temperature cycling at a specified rate; select the model against the required ramp, dwell time and specimen load.

Define in your protocol

  • Upper/lower temperatures, rate, dwell time and cycles
  • Specimen temperature response and measurement points

Confirm for this chamber

  • Ramp performance for the actual load and heat dissipation
  • Distinction between air temperature and specimen temperature

Official standard reference ↗

Related Industry Requirements

The listed methods address temperature exposure. They do not establish turntable accuracy, rate performance or calibration uncertainty. The applicable edition, test severity, load and monitoring requirements are confirmed for the ordered configuration. Standard references do not imply complete product certification.

Technical Specifications

Current standard specifications
Model / platformCTT-SC7512 · 2 / 5°C/minCTT-SC7525 · 2 / 5°C/minCTT-SC7550 · 2 / 5°C/minCTT-SC75100 · 2 / 5°C/min
Nominal workspaceL1252505001000
Interior dimensionsW × H × D (mm)475 × 560 × 475600 × 700 × 600750 × 900 × 750950 × 1150 × 950
One-piece system exterior dimensionsW × H × D (mm)675 × 1700 × (1200+700)800 × 1750 × (1350+700)950 × 1750 × (1600+700)1150 × (2000+150) × (1850+700)
Split-system chamber dimensionsW × H × D (mm)675 × 1450 × 1200800 × 1650 × 1350950 × 1750 × 16001150 × (2000+150) × 1850
Separate refrigeration unit dimensionsW × H × D (mm)700 × 1700 × 700700 × 1700 × 1100700 × 1700 × 1100700 × 1700 × 1200
Observation windowW × H (mm)315 × 315350 × 440440 × 600440 × 600
Cable port diametermm100
Unit weight · 2 / 5°C/minkg280/300380/410530/630740/770
Temperature range°C−75°C to +150°C
Temperature fluctuation°C≤±0.5°C; empty with platform installed, steady state
Temperature uniformity°C≤2.0°C; empty with platform installed, steady state
Temperature deviation°C≤±2.0°C; empty with platform installed, steady state
Average heating / cooling rate°C/min2 / 5°C/min; average; empty with platform; +85°C → −55°C
Platform diameter limitmm≤350≤500≤650≤850
Installed power · 2 / 5°C/minkW6/8.58.5/1212/1515/22
Supply circuit breaker · 2 / 5°C/minA16/2525/3232/4040/63
Cooling-water requirement2/2.4 m³/h for 2 / 5°C/min configurations2.4/3.5 m³/h for 2 / 5°C/min configurations3.5/4.5 m³/h for 2 / 5°C/min configurations4.5/6.5 m³/h for 2 / 5°C/min configurations
Cooling-water connectionDN25DN25DN25DN25/DN40
Temperature sensorClass A PT100
Communication interfacesOne RS232 interface; one RJ45 Ethernet interface; one USB data interface; RS485 optional
Paired power, breaker, weight and water-flow values are for 2 / 5°C/min configurations. Split-system chamber dimensions and separate refrigeration-unit dimensions are listed independently. Model suffixes are specified for the selected rate and cooling configuration.
Earlier model configurationsSelect an existing model reference below when enquiring about a previous configuration. Availability and the ordered configuration are confirmed in the quotation.
Main Technical Specifications
Model (Single-Axis Turntable Chamber)Model (Single-Axis Turntable Chamber)CTT-SC7506
Model (Vibration-Isolation Platform Chamber)Model (Vibration-Isolation Platform Chamber)CTT-ST7506
Nominal VolumeL60
Interior DimensionsW×H×D(mm)450×370×370
Exterior DimensionsW×H×D(mm)650×1300×1150
Temperature Range℃-75°C to +150°C (customizable)
Temperature Performance℃Fluctuation: ±0.5°C; uniformity: 2.0°C (no-load test with platform installed, within the effective uniform zone); temperature deviation: ±2.0°C
Heating / Cooling Rate℃/min-02: ≥2.0°C/min; -05: ≥5.0°C/min (no-load test with platform installed; test range +85°C to -55°C)
Operating Noise (-02 / -05)dB(A)62/64
Installed Power (-02 / -05)kW2.5/4.0
Equipment Weightkg250
Turntable Diametermm≤Φ300
Connection Method–Flexible sealed shaft opening between the chamber floor and the turntable / platform
Vibration-Isolation Platform Configuration–Granite vibration-isolation platform; the granite top and base are connected by height-adjustable supports. Available only on the CTT-ST series.
Power Supply–380 V ±10% AC, 50 Hz, 3L + N + PE; dedicated supply with separate circuit breaker and reliable grounding
Configuration — Observation WindowmmW260×H260
Configuration — Cable Ports–One cable port on each side of the chamber, supplied with two silicone plugs
Configuration — Controller–High-brightness Chinese-language touchscreen with imported programmable controller
ENGINEERING CONSULTATION

Turn your requirement into a buildable chamber specification.

Share the essentials. Our application engineers will help define a practical, reliable system.

Application and standards reviewPractical chamber configurationInitial response within 1 business day