Primary Applications
- Temperature exposure during single-axis turntable testing of inertial sensors and navigation assemblies.
- Evaluation of temperature-dependent measurement drift, alignment and functional stability.
- Conditioned static measurements and calibration using the specified external instruments.
- Integrated motion and climatic test development for aerospace, electronics and precision equipment.
Product Features
- Chamber interface engineered around a single-axis turntable, specimen fixture and service connections.
- Workspace geometry, access and clearances are selected for the actual turntable installation.
- Temperature range, rate and load conditions are stated for the listed chamber model.
- Programmable temperature control with communication interfaces for system integration.
- Motion performance and measurement uncertainty are specified for the complete turntable and instrumentation system.
Applicable Test Standards
Select the exposure method, then confirm the test profile and specimen load for your application.
Cold exposure of components and equipment.
Define in your protocol
- Lower temperature, tolerances and exposure duration
- Specimen operating state, stabilization and recovery
Confirm for this chamber
- Minimum temperature for the selected model and load
- Usable test space, sensors and loaded uniformity
Dry-heat exposure of components and equipment.
Define in your protocol
- Upper temperature, exposure duration and test sequence
- Specimen heat dissipation, operating state and recovery
Confirm for this chamber
- Maximum temperature and actual specimen heat release
- Loaded uniformity, airflow and independent monitoring
Temperature cycling at a specified rate; select the model against the required ramp, dwell time and specimen load.
Define in your protocol
- Upper/lower temperatures, rate, dwell time and cycles
- Specimen temperature response and measurement points
Confirm for this chamber
- Ramp performance for the actual load and heat dissipation
- Distinction between air temperature and specimen temperature
Related Industry Requirements
MIL-STD-810H + Change 1
Environmentally tailored profiles
Select the method and severities for the intended service environment. Confirm the applicable revision, specimen operating state, monitoring locations and loaded test performance.
ISO 16750-4:2023
Climatic loads for vehicle electronics
Select the climatic program for the mounting location and intended operating state. Confirm specimen supply, heat dissipation, tolerances and functional checks.
IEC 60068 series / laboratory protocols
Document the actual exposure method
Define the relevant method or laboratory procedure, specimen load, measurement locations, calibration and repeatability criteria. Equipment selection follows the required profile and documented uncertainty.
The listed methods address temperature exposure. They do not establish turntable accuracy, rate performance or calibration uncertainty. The applicable edition, test severity, load and monitoring requirements are confirmed for the ordered configuration. Standard references do not imply complete product certification.
Technical Specifications
| Current standard specifications | |||||
|---|---|---|---|---|---|
| Model / platform | CTT-SC7512 · 2 / 5°C/min | CTT-SC7525 · 2 / 5°C/min | CTT-SC7550 · 2 / 5°C/min | CTT-SC75100 · 2 / 5°C/min | |
| Nominal workspace | L | 125 | 250 | 500 | 1000 |
| Interior dimensions | W × H × D (mm) | 475 × 560 × 475 | 600 × 700 × 600 | 750 × 900 × 750 | 950 × 1150 × 950 |
| One-piece system exterior dimensions | W × H × D (mm) | 675 × 1700 × (1200+700) | 800 × 1750 × (1350+700) | 950 × 1750 × (1600+700) | 1150 × (2000+150) × (1850+700) |
| Split-system chamber dimensions | W × H × D (mm) | 675 × 1450 × 1200 | 800 × 1650 × 1350 | 950 × 1750 × 1600 | 1150 × (2000+150) × 1850 |
| Separate refrigeration unit dimensions | W × H × D (mm) | 700 × 1700 × 700 | 700 × 1700 × 1100 | 700 × 1700 × 1100 | 700 × 1700 × 1200 |
| Observation window | W × H (mm) | 315 × 315 | 350 × 440 | 440 × 600 | 440 × 600 |
| Cable port diameter | mm | 100 | |||
| Unit weight · 2 / 5°C/min | kg | 280/300 | 380/410 | 530/630 | 740/770 |
| Temperature range | °C | −75°C to +150°C | |||
| Temperature fluctuation | °C | ≤±0.5°C; empty with platform installed, steady state | |||
| Temperature uniformity | °C | ≤2.0°C; empty with platform installed, steady state | |||
| Temperature deviation | °C | ≤±2.0°C; empty with platform installed, steady state | |||
| Average heating / cooling rate | °C/min | 2 / 5°C/min; average; empty with platform; +85°C → −55°C | |||
| Platform diameter limit | mm | ≤350 | ≤500 | ≤650 | ≤850 |
| Installed power · 2 / 5°C/min | kW | 6/8.5 | 8.5/12 | 12/15 | 15/22 |
| Supply circuit breaker · 2 / 5°C/min | A | 16/25 | 25/32 | 32/40 | 40/63 |
| Cooling-water requirement | 2/2.4 m³/h for 2 / 5°C/min configurations | 2.4/3.5 m³/h for 2 / 5°C/min configurations | 3.5/4.5 m³/h for 2 / 5°C/min configurations | 4.5/6.5 m³/h for 2 / 5°C/min configurations | |
| Cooling-water connection | DN25 | DN25 | DN25 | DN25/DN40 | |
| Temperature sensor | Class A PT100 | ||||
| Communication interfaces | One RS232 interface; one RJ45 Ethernet interface; one USB data interface; RS485 optional | ||||
Earlier model configurations
Select an existing model reference below when enquiring about a previous configuration. Availability and the ordered configuration are confirmed in the quotation.| Main Technical Specifications | ||
|---|---|---|
| Model (Single-Axis Turntable Chamber) | Model (Single-Axis Turntable Chamber) | CTT-SC7506 |
| Model (Vibration-Isolation Platform Chamber) | Model (Vibration-Isolation Platform Chamber) | CTT-ST7506 |
| Nominal Volume | L | 60 |
| Interior Dimensions | W×H×D(mm) | 450×370×370 |
| Exterior Dimensions | W×H×D(mm) | 650×1300×1150 |
| Temperature Range | ℃ | -75°C to +150°C (customizable) |
| Temperature Performance | ℃ | Fluctuation: ±0.5°C; uniformity: 2.0°C (no-load test with platform installed, within the effective uniform zone); temperature deviation: ±2.0°C |
| Heating / Cooling Rate | ℃/min | -02: ≥2.0°C/min; -05: ≥5.0°C/min (no-load test with platform installed; test range +85°C to -55°C) |
| Operating Noise (-02 / -05) | dB(A) | 62/64 |
| Installed Power (-02 / -05) | kW | 2.5/4.0 |
| Equipment Weight | kg | 250 |
| Turntable Diameter | mm | ≤Φ300 |
| Connection Method | – | Flexible sealed shaft opening between the chamber floor and the turntable / platform |
| Vibration-Isolation Platform Configuration | – | Granite vibration-isolation platform; the granite top and base are connected by height-adjustable supports. Available only on the CTT-ST series. |
| Power Supply | – | 380 V ±10% AC, 50 Hz, 3L + N + PE; dedicated supply with separate circuit breaker and reliable grounding |
| Configuration — Observation Window | mm | W260×H260 |
| Configuration — Cable Ports | – | One cable port on each side of the chamber, supplied with two silicone plugs |
| Configuration — Controller | – | High-brightness Chinese-language touchscreen with imported programmable controller |