Primary Applications
- Cold and dry-heat exposure of electronic components, assemblies and materials.
- Temperature cycling to evaluate functional drift, material compatibility and assembly reliability.
- Development validation, production screening and failure investigation within the selected model’s capability.
- Conditioned storage and preconditioning before electrical, mechanical or dimensional measurements.
Product Features
- Compact chamber format for component and small-assembly temperature testing.
- Model-specific temperature limits, heating/cooling rates and test conditions shown in the technical specification matrix.
- An operator interface for setting and monitoring the required temperature exposure.
- Equipment and specimen protection functions, with the final configuration matched to the application.
Applicable Test Standards
Select the exposure method, then confirm the test profile and specimen load for your application.
Cold exposure of components and equipment.
Define in your protocol
- Lower temperature, tolerances and exposure duration
- Specimen operating state, stabilization and recovery
Confirm for this chamber
- Minimum temperature for the selected model and load
- Usable test space, sensors and loaded uniformity
Dry-heat exposure of components and equipment.
Define in your protocol
- Upper temperature, exposure duration and test sequence
- Specimen heat dissipation, operating state and recovery
Confirm for this chamber
- Maximum temperature and actual specimen heat release
- Loaded uniformity, airflow and independent monitoring
Temperature cycling at a specified rate; select the model against the required ramp, dwell time and specimen load.
Define in your protocol
- Upper/lower temperatures, rate, dwell time and cycles
- Specimen temperature response and measurement points
Confirm for this chamber
- Ramp performance for the actual load and heat dissipation
- Distinction between air temperature and specimen temperature
Related Industry Requirements
ISO 16750-4:2023
Climatic loads for vehicle electronics
Select the climatic program for the mounting location and intended operating state. Confirm specimen supply, heat dissipation, tolerances and functional checks.
JESD22 / AEC-Q100
Device-specific methods and controlled revision
Select the applicable controlled method, exposure medium, specimen measurements and acceptance criteria. Electrical bias and complete device qualification require the specified additional apparatus and procedures.
IEC 60068 series / laboratory protocols
Document the actual exposure method
Define the relevant method or laboratory procedure, specimen load, measurement locations, calibration and repeatability criteria. Equipment selection follows the required profile and documented uncertainty.
The applicable edition, test severity, load and monitoring requirements are confirmed for the ordered configuration. Standard references do not imply complete product certification.
Technical Specifications
| Current standard specifications | ||||
|---|---|---|---|---|
| Model / platform | CTE-SG4506-02F | CTE-SG7506-02F | CTE-SE7506-05F | |
| Nominal workspace | L | 60 | ||
| Interior dimensions | W × H × D (mm) | 450 × 370 × 370 | ||
| Exterior dimensions | W × H × D (mm) | 610 × 1350 × 807 | ||
| Cable port diameter | mm | 100 | ||
| Unit weight | kg | 250 | ||
| Temperature range | °C | -45 to +150°C | -75 to +150°C | -75 to +150°C |
| Temperature fluctuation | °C | ≤±0.5°C; empty, steady state | ||
| Temperature uniformity | °C | ≤2.0°C below +60°C; ≤3.0°C at or above +60°C; empty, steady state | ||
| Temperature deviation | °C | ≤±2.0°C; empty, steady state | ||
| Average heating / cooling rate | °C/min | ≥2 | ≥2 | ≥5 |
| Rate-test temperature interval | °C | +85°C → -40°C | +85°C → -55°C | +85°C → -55°C |
| Rate-test aluminium load | kg | Empty chamber | ||
| Power supply | 220 V, 50 Hz | 220 V, 50 Hz | Specified for the selected electrical / cooling configuration | |
| Installed power | kW | 3.3 | 3.3 | 4 |
| Supply circuit breaker | A | 16 | ||
| Cooling-water requirement | Air-cooled | |||
| Temperature sensor | Class A PT100 | |||
| Refrigeration system | Single-compressor refrigeration system | |||
| Communication interfaces | One RS232 interface; one RJ45 Ethernet interface; one USB data interface; RS485 optional | |||
Average ramp rates are measured at the air outlet over the stated interval with the stated specimen load. A non-self-heating aluminium load is a test condition.
Construction and configuration
| Controller | 7-inch high-brightness Chinese-language touchscreen with imported programmable controller |
|---|
Earlier model configurations
Select an existing model reference below when enquiring about a previous configuration. Availability and the ordered configuration are confirmed in the quotation.
| Main Technical Specifications | |||||
|---|---|---|---|---|---|
| Model | Model | CTE-SG4506-02F/W | CTE-SG7506-02F/W | CTE-SE7506-05F/W | CTE-SE7506-10F/W |
| Nominal Volume | L | 60 | 60 | 60 | 60 |
| Interior Dimensions | W×H×D(mm) | 450×370×370 | 450×370×370 | 450×370×370 | 450×370×370 |
| Exterior Dimensions | W×H×D(mm) | 610×1250×807 | 610×1250×807 | 610×1350×807 | 610×1350×807 |
| Temperature Range | ℃ | -45~+150 | -75~+150 | -75~+150 | -75~+150 |
| Temperature Performance | ℃ | Uniformity: ≤2.0°C below 60°C and ≤3.0°C at or above 60°C; fluctuation: ≤±0.5°C; deviation: ≤±2.0°C | Uniformity: ≤2.0°C below 60°C and ≤3.0°C at or above 60°C; fluctuation: ≤±0.5°C; deviation: ≤±2.0°C | Uniformity: ≤2.0°C below 60°C and ≤3.0°C at or above 60°C; fluctuation: ≤±0.5°C; deviation: ≤±2.0°C | Uniformity: ≤2.0°C below 60°C and ≤3.0°C at or above 60°C; fluctuation: ≤±0.5°C; deviation: ≤±2.0°C |
| Heating / Cooling Rate (No Load) | ℃/min | ≥2.0(+85℃~-40℃) | ≥2.0(+85℃~-55℃) | ≥5.0(+85℃~-55℃) | ≥10.0(+85℃~-40℃) |
| Linear Temperature Control (Optional, No Load) | ℃/min | 0.5~1.0(+60℃~-25℃) | 0.5~1.0(+60℃~-40℃) | 0.5~2.0(+60℃~-40℃) | 0.5~5.0(+60℃~-40℃) |
| Operating Noise | dB(A) | 52~62 | 52~62 | 52~65 | 52~68 |
| Installed Power | kW | 2.0 / 3.5 | 2.8 / 4.5 | 4.5 / 6.0 | 7.5 / 9.5 |
| Equipment Weight | kg | 225/275 | 230/280 | 250/300 | 280/330 |
| Power Supply | – | 220V±10%VAC | 220V±10%VAC | 380V±10%VAC | 380V±10%VAC |
| Product Configuration | Observation Window | One coated, heated, anti-frost insulated-glass observation window, W260 × H260 mm, with LED lighting | One coated, heated, anti-frost insulated-glass observation window, W260 × H260 mm, with LED lighting | One coated, heated, anti-frost insulated-glass observation window, W260 × H260 mm, with LED lighting | One coated, heated, anti-frost insulated-glass observation window, W260 × H260 mm, with LED lighting |
| Product Configuration | Cable Ports | One Ø80 mm cable port on each side, supplied with two silicone plugs | One Ø80 mm cable port on each side, supplied with two silicone plugs | One Ø80 mm cable port on each side, supplied with two silicone plugs | One Ø80 mm cable port on each side, supplied with two silicone plugs |
| Product Configuration | Specimen Shelf | Height-adjustable stainless steel; quantity: 1; load capacity: 5 kg per shelf | Height-adjustable stainless steel; quantity: 1; load capacity: 5 kg per shelf | Height-adjustable stainless steel; quantity: 1; load capacity: 5 kg per shelf | Height-adjustable stainless steel; quantity: 1; load capacity: 5 kg per shelf |
| Product Configuration | Controller | 7-inch high-brightness Chinese-language touchscreen with imported programmable controller | 7-inch high-brightness Chinese-language touchscreen with imported programmable controller | 7-inch high-brightness Chinese-language touchscreen with imported programmable controller | 7-inch high-brightness Chinese-language touchscreen with imported programmable controller |